Get Empowered: Review Window and the Processing Method | Tip #101, ACQUITY QDa MS Data

By December 26, 2018


Tip #101: Working with ACQUITY QDa MS data in Empower Software (Part 2)

Welcome back to Get Empowered! In the last Empower tip-of-the-week post for Empower Chromatography Data Software, we learned how to view the base mass and mass spectra for the peaks in a chromatogram (Tip #100).

This week, we will learn how to work with 2D ACQUITY QDa MS data (SIR which is ‘Single Ion Recording’ of data at a particular mass) and create a basic LC Processing Method for quantitative analysis. Which tabs of the Processing Method do I need to use?

Let’s get started.

Step 1

Bring the 2D channels for standards and samples into Review. In this case, we are quantitating Aspartame, an artificial sweetener used in soft drinks. The mass is 295.2 (figure 1).

Empower Tips: 2D channels for standards and samples into Review | Figure 1

Figure 1

Step 2

Once the peak detection and integration parameters have been set, you may want to include ‘smoothing’ depending on the Signal to Noise ratio. In this case, it is not needed (figure 2).

Empower Tips: Include ‘smoothing’ depending on the Signal to Noise ratio | Figure 2

Figure 2

Step 3

On the Components tab, enter the ‘peak name’ and any other pertinent information. In this case, the method calls for a quadratic fit for the calibration curve (figure 3).

Empower Tips: Components tab | Figure 3

Figure 3

Step 4

Optionally turn on System Suitability to have Signal to Noise plate count and so forth calculated. Save the method and test on your standards and samples (figure 4).

Empower Tips: Save the method and test on your standards and samples | Figure 4

Figure 4

Step 5

Once the Sample Set has been processed, take the Result Set into Review, and view the calibration curve (figure 5).

Empower Tips: bring result Set into review and view the calibration curve | Figure 5

Figure 5

Step 6

Review the samples to view the amount (or concentration) of Aspartame (figure 6).

Empower Tips: Review the sample | Figure 6

Figure 6

As you can see it is no different that working with 2D UV data.

It’s that easy!

 

Final notes: 

  • This procedure can be followed using the QuickStart or Pro interface.
  • ACQUITY QDa Mass Detector is compatible with the Alliance HPLC system.

 

Please rate this Empower Tip of the Week


Next week in Empower Tip #102– We will continue this series on working with ACQUITY QDa MS data in Empower (Part 3).

More resources

 

Do you want Empower Tips sent to your inbox every week?

Subscribe Now

Questions? Tips of your own? Let us know!

 


Categories: Empower Tips