Get Empowered: Tip #255 | Peak Purity (Part 5)
Tip #255: Peak Purity Parameters for PDA Data
Welcome back to Get Empowered! In this week’s tip for Empower Chromatography Data System (CDS), we will discuss Peak Purity parameters for Photodiode Array (PDA) data. (In last week’s tip, we discussed the Spectral Contrast parameters for PDA data, Tip #254).
Let’s get started.
Step 1
On the ‘Purity’ tab of the Processing Method there are several Purity Parameters. To begin, check the ‘Purity Enabled’ box to have Peak Purity calculated. The Active Peak Region (%) defaults to 100.0. This means Peak Purity will be calculated using all spectra from ‘peak liftoff’ to ‘peak touchdown’ (figure 1).

Figure 1
Step 2
If the baseline is noisy, you may want to enter a value less than 100.0 to ensure Empower does not use noisy baseline spectra to calculate Peak Purity (figures 2 & 3).

Figure 2

Figure 3
Step 3
The Threshold Angle can be calculated in a variety of ways. Typically, it is the sum of the Noise Angle (which is based on the noise interval selected) plus the Solvent Angle. It is recommended to try AutoThreshold first. AutoThreshold determines the Solvent Angle for each peak in the chromatogram. The Solvent Angle is automatically calculated as a function of the Maximum Spectral Absorbance (MSA) of a peak. MSA is the maximum value of baseline-corrected spectral absorbance for a peak. AutoThreshold does need to be validated and there are some limitations (figure 4):

Figure 4
- Make six injections of the standard (figure 5).

Figure 5
- Create a PDA Processing Method and follow Steps 1 & 2 above.
- Determine the ‘Noise Interval’ as described in Tip #254.
- Set the Threshold Angle to ‘AutoThreshold’.
- Process the six injections. If the ‘Purity Angle’ is less than the ‘Purity Threshold’ for all peaks in all injections, then AutoThreshold is working (figure 6).

Figure 6
- You can use the Threshold Angle obtained from AutoThreshold for Peak Purity testing on any unknown, as long as its maximum spectral absorbance is less than 1.0 AU and less than five times the maximum spectral absorbance of the standard used to develop the method.
Step 4
If the Purity Angle is greater than the Purity Threshold for any of the peaks, in any of the standard injections, you need to determine the Solvent Angle separately. That will be described in next week’s Tip #256.
It’s that easy!
Final Note: This can be done with either the Pro or QuickStart interface.
Please rate this Empower Tip of the Week
Next time in Empower Tip #256– we will continue our series on Peak Purity using both PDA and QDa Detectors.
JOIN US at [inform] 2022 | May 9-12, 2022 · Nashville, TN USA
Join the Waters Informatics community to discuss the latest forward-thinking digital innovations and technologies to elevate quality and improve laboratory effectiveness.
More Empower Resources:
- Get the latest Empower projects, options, and more. Check out the New Waters Marketplace.
- Want to learn more? Check out our Empower training courses and eLearning.
- Minimize your lab downtime and protect your business against potentially expensive disruptions with the Empower Enterprise Software Extended Critical Support Plan with 24 x 7 x 365 coverage – Learn more
- Build your technical skills and professional network. Share, ask, and interact with peers from around the globe in dynamic discussions by joining our Informatics Community Forum.
Do you want Empower Tips sent to your inbox every week?
Questions? Tips of your own? Let us know! |