Get Empowered: Tip #243 | Method Versioning (Part 2)

By September 15, 2021


Tip #243: Method Versioning: Sample Set Methods

Welcome back to Get Empowered! In the last Empower tip-of-the-week blog for Empower Chromatography Data System, we began a new series on Method Versioning (Tip #242).

In this week’s tip, we continue the series on Method Versioning, focusing on Sample Set Methods.

Let’s get started.

Altering a Running Sample Set During Acquisition

Step 1

Sample Set Method ‘23August’, ‘Method Version 1’ is loaded into Run Samples and Run (figure 1).

Figure 1

 Step 2

The Running Sample Set is altered; the Sample Set Method ‘23August’ is saved resulting in the creation of ‘Method Version 2’ which becomes the current version (figure 2).

Figure 2

Altering a Sample Set After Acquisition

Step 3

An Acquired Sample Set is altered; the ‘Sample Name’ for the first sample is changed (figure 3).

Figure 3

Step 4

The current version of the Sample Set Method (Method Version 2) remains unchanged; however, ‘Method Version 3’ of the method is created as a result of the change carried out in Step 3. Method Version 3 can be accessed from Method Properties (figure 4).

Figure 4

Step 5

Method Differences displays the change in the ‘Sample Name’ for the first sample between Method Version 2 and Method Version 3 (figure 5).

Figure 5

Acquiring a Sample Set Using a Non-Current Version of the Sample Set Method

Step 6

In the ‘Run Samples’ window, click ‘Load’ and select ‘Re-inject samples from a previously run sample set’, then click ‘OK’ (figure 6).

Figure 6

Step 7

Select the Sample Set from the project window, click ‘Load into Run Samples’. This Sample Set was acquired with Sample Set Method SSM 1, Method Version 2 (figure 7).

Figure 7

Step 8

The samples are loaded into Run Samples and the Sample Set Method associated with this Sample Set is used to acquire the data (figure 8).

Figure 8

Making a Sample Set’s Non-Current Version of the Sample Set Method into the Current Version

Step 9

Repeat Step 6 and Step 7 above. From the File menu in Run Samples, select ‘Save Sample Set Method As’ (figure 9).

Figure 9

Step 10

Leave the name of the method the same, click ‘Save’. You will be prompted: ‘Do you want to overwrite the existing sample set method with this name?’, click ‘Yes’ (figure 10).

Figure 10

Step 11

Run the Sample Set Method (figure 11).

Figure 11

Step 12

The Sample Set Method, SSM 2 is now at Method Version 2 (figure 12). (NOTE: You may recall that SSM 2 was previously at Method Version 1.

Figure 12

 

It’s that easy!

 

Final Note: This can be done with either the Pro or QuickStart interface.

 

Please rate this Empower Tip of the Week

empower


Next week in Empower Tip #243: We continue the series on Method versioning and discuss Sample Set Methods.

 

empower

JOIN US for an Empower: Ask Me Anything Webinar on September 22, 2021

Neil will answer your questions about streamlining your Peak Purity workflow with Empower using the PDA and QDa Mass Detector.

Learn More and Register 

 