Get Empowered: Review Window and the Processing Method | Tip #70, System Suitability

By May 23, 2018


Tip #70: Working with the System Suitability Option (Part 3)

Welcome back to Get Empowered! In our last tip-of-the-week blog for Empower Software, we learned about how Peak to Peak Noise is calculated in Empower (Tip 69).

In this week’s tip, we are going to learn how Empower calculates Baseline Noise and Drift (part 3).

If you are not in a regulated laboratory environment, you probably want to calculate Signal to Noise without the USP, EP, or JP formula (Tip #28). One of the options for noise is Baseline Noise. This is the original option that was available with Millennium chromatography software in the 1990s.

Let me show you how it is done.

Step 1

We begin by bringing a Sample Set into Review, open a Processing Method, and Integrate/Calibrate or Integrate/Quantitate a channel (Figure 1).

Empower Tips: Integrate/Calibrate or Integrate/Quantitate a Channel | Figure 1

Figure 1

Step 2

Go to the Processing Method window, click the Suitability tab, and select Baseline Noise from the Noise Value for s/n drop-down list (Figure 2).

Empower Tips: Select Baseline Noise from the Noise Value for s/n drop down list | Figure 2

Figure 2

Step 3

The Baseline Noise minimum is either a minimum time interval or minimum number of data points used to calculate baseline noise (Figure 3).

Empower Tips: The Baseline Noise minimum | Figure 3

Figure 3

Step 4

So, how does Empower calculate Baseline Noise? The software calculates noise based on the maximum signal change over a 30-second interval. The reported noise value is an average of two 30-second intervals, as determined by the parameter % Run Time Over Which to Average. To determine the 30-second intervals, the software begins counting from the Baseline Start time and the Baseline End time toward the middle of the chromatogram.

In figure 4, we have a 10-minute run time. 0.5% of 10 minutes is 0.5 minute. Therefore, region 1 is 1 to 1.5 minutes and region 2 is 8.5 to 9 minutes. Be sure to pick areas of the baseline that only contain noise and not peaks, otherwise you will have a high noise value reported.

Empower Tip: Calculating Baseline Noise in Empower | Figure 4

Figure 4

Step 5

In our chromatogram, we have a 15-minute run time, therefore, if we use 3.5 for the % Run Time Over Which to Average, that will give us 30-second intervals. We set the Baseline Start to 0 and the Baseline End to 15 (Figure 5).

Empower Tips: Setting Baseline Start and End | Figure 5

Figure 5

Step 6

Return to the Main Window and Integrate/Calibrate or Integrate/Quantitate. The Signal to Noise is calculated using Baseline Noise for each integrated peak (Figure 6).

Empower Tips: Signal to Noise calculation | Figure 6

Figure 6

Step 7

Go to the Results window and we can see the Baseline Noise, Baseline Drift, and Scale to microvolts fields in the Chromatogram Result table (Figure 7). The Baseline Drift is simply the signal at Baseline End minus the signal at Baseline Start.

Empower Tips: View Chromatogram Result table | Figure 7

Figure 7

 

It’s that easy!

Final Notes:

  1. Peak height is in microvolts and baseline noise is in millivolts, therefore, Empower uses the scale to microvolts factor to convert the baseline noise into microvolts.
  2. This procedure can be followed using the QuickStart or Pro interface.

 

Please rate this Empower Tip of the Week


 

Next week in Empower Tip #71 – Working with the System Suitability Option in Empower (part 4)

 

 More resources:

 

Do you want Empower Tips sent to your inbox every week?

Subscribe Now

Questions? Tips of your own? Let us know!

 

< Previous
Next >

Categories: Empower Tips