Get Empowered: Review Window and the Processing Method | Tip #31

By August 23, 2017


Tip #31: Calculating signal-to-noise ratio (Part 4)

Welcome back to Get Empowered! In this mini-series for our tip-of-the-week blog on using Empower Software, we’re exploring how to calculate signal-to-noise (S/N) ratio. This tip will demonstrate a fourth option, using custom calculations.

If you work in a non-regulated laboratory and do not have the System Suitability option for Empower Software, you may still need to calculate signal-to-noise ratio. We can do this by creating a simple custom calculation. Sound difficult? Well it isn’t. Let me show you how.

Step 1

Let’s start by creating the custom calculation. This is done in the properties of your project. Go to the Configuration Manager window by clicking Configure the System from the Pro Interface (Figure 1).

Empower Tips: Configure System | Figure 1

Figure 1

Step 2

Right-click on the row number for your project and select Properties (Figure 2).

Empower Tips: Select properties | Figure 2

Figure 2

Step 3

Click the Custom Fields tab and then click the New button to launch the New Custom Field Wizard (Figure 3).

Empower Tips: New custom field wizard | Figure 3

Figure 3

Step 4

On the Data and Type Selection screen, select Peak for the Field Type and Real for the data type (Figure 4). A peak custom field is a calculation that can be applied to every peak in the chromatogram. Click Next.

Empower Tips: Select Peak for the field type and Real for the data type | Figure 4

Figure 4

Step 5

Do not change anything on the Source Selection screen (Figure 5). Click Next.

Empower Tips: Source Selection screen | Figure 5

Figure 5

Step 6

We will enter the formula on the Formula Entry screen. Signal-to-noise ratio is peak height divided by detector noise. In the Fields list scroll to the right and double-click on Height (Figure 6).

Empower Tips: Enter formula/Formula entry screen | Figure 6

Figure 6

Step 7

From the Operations list, double-click the forward slash, which is the division symbol (Figure 7).

Empower Tips: Operations list/double click the forward slash | Figure 7

Figure 7

Step 8

Height is in microvolts and detector noise is in plot units, so we will need to correct for the difference in units. In the denominator enter a left parenthesis (Figure 8).

Empower Tips: Denominator enter a left parenthesis | Figure 8

Figure 8

Step 9

In the Fields list, locate Detector Noise and double-click to add it to the formula. Then double-click the forward slash (Figure 9).

Empower Tips: Detector noise/double click to add it to the formula | Figure 9

Figure 9

Step 10

In the Fields list, locate Scale to uV and double- click to add it to the formula. The Scale to uV will correct for the difference in units. Then add the right parenthesis (Figure 10). Click Next.

Empower Tips: Scale to uV / double click to add it to the formula | Figure 10

Figure 10

Step 11

Set the desired precision on the Numeric Parameters screen and click Next (figure 11).

Empower Tips: Set the desired precision/Numeric Parameters screen | Figure 11

Figure 11

Step 12

Enter a name on the Name Entry screen and click Finish (Figure 12). Note: you cannot use spaces in the name but could do shift underscore.

Empower Tips: Enter a name/Name Entry screen | Figure 12

Figure 12

Step 13

Click OK to the warning message (Figure 13).

Empower Tips: Warning message | Figure 13

Figure 13

Step 14

Now let’s use the custom field. Open the project and bring raw data (Sample Set, Injection or Channel) into review and open the Processing Method. Click the Processing Method tool (Figure 14).

Empower Tips: custom field/Processing method tool | Figure 14

Figure 14

Step 15

Click the Noise and Drift tab (Figure 15).

Empower Tips: Noise & drift tab | Figure 15

Figure 15

Step 16

Check the box for Calculate Detector Noise and Drift and enter a time interval that is free of peaks for calculating noise (Figure 16). This time region must be equal to or greater than the Segment Width and must contain at least 60 data points. The Segment Width is used for calculating average noise and drift values and it must contain at least 30 data points.

Empower Tips: Calculate Detector Noise and Drift | Figure 16

Figure 16

Step 17

Return to the Main Window and click the Integrate tool (Figure 17).

Empower Tips: Main window/Integrate tool | Figure 17

Figure 17

Step 18

If signal-to-noise ratio is not visible we can add it to the Peaks table. Right-click on the Peaks table and select Table Properties (Figure 18).

Empower Tips: Add signal-to-noise to the Peaks table | Figure 18

Figure 18

Step 19

On the Columns tab, find signal-to-noise ratio and remove the check mark to make it visible in the table (Figure 19). Click OK.

Empower Tips: Check mark to make signal-to-noise visible in the table | Figure 19

Figure 19

Step 20

The signal-to-noise ratio is displayed for every integrated peak (Figure 20).

Empower Tips: Signal-to-noise ratio displayed for every integrated peak | Figure 20

Figure 20

It’s that easy!

I can calculate signal-to-noise ratio in Empower! #EmpowerTips Click To Tweet

Final note:

  1. This procedure can be followed using the QuickStart interface. In step 1 described above, select Configuration from the Manage menu.

 


Please rate this Empower Tip of the Week


 

Next week in Empower Tip #32: Overlaying 2D Chromatograms from Different 3D PDA Channels

 