Get Empowered: Review Window and the Processing Method | Tip #115, ACQUITY QDa MS Data

By April 3, 2019


Tip #115: Working with ACQUITY QDa MS data in Empower Software (Part 15)

Welcome back to Get Empowered! In the last Empower tip-of-the-week post for Empower Chromatography Data Software, we learned how to use a MS Spectra Library to identify spectra from unknown peaks via the automated technique (Tip  #114).

Working with ACQUITY QDa MS data in Empower:

  • #100: How to view the base mass and mass spectra for the peaks in a chromatogram
  • #101: How to work with 2D ACQUITY QDa MS data
  • #103: Viewing different spectral views in the Mass Analysis window
  • #104: Extracting chromatograms and spectra while working in the Mass Analysis window
  • #105: Why it is important to align chromatograms when collecting data from both a PDA and a ACQUITY QDa mass detector
  • #106: How to optimize sensitivity by extracting a single mass rather than working with a TIC plot
  • #107: How to smooth the chromatogram using the Processing Method
  • #108: How to smooth the chromatogram using a Derived Channel in the Method Set
  • #109: How to track peaks using the ACQUITY QDa mass detector
  • #110: Using CODA (component detection algorithm) as a tool when scouting for peaks
  • #111: Using MS 3D subtraction to remove background interference
  • #112: How to create a MS Spectra Library from data collected with the ACQUITY QDa mass detector
  • #113: How to use a MS Spectra Library to identify spectra from unknown peaks via the manual technique
  • #114: How to use a MS Spectra Library to identify spectra from unknown peaks via the automated technique

This week, I would like to answer a popular question I received from several readers.

Q: If I collect multiple 3D scans with my ACQUITY QDa, how do I display spectra from an MS scan along with my UV spectra in the Mass Analysis window?

A: It can be done in 4 simple steps. Let me show you how.

Let’s get started.

Step 1

We start in the Main Window of Review. View the PDA scan and the two MS scans listed in the 3D Channels table (figure 1).

Empower Tips: View the PDA scan and the two MS scans | Figure 1

Figure 1

Step 2

In the Mass Analysis Window, we see the UV spectra and MS spectra from the ACQUITY QDa 1 scan (figure 2).

Empower tips: View the UV spectra and MS spectra | Figure 2

Figure 2

Step 3

Go to the Processing Method and select the MS 3D Channel tab. Towards the bottom of the window, select which MS scan you would like to use on a peak by peak basis from the 3D Channel Name (Description) field (figure 3).

Empower Tips: Select the MS scan peak by peak basis | Figure 3

Figure 3

Step 4

Reapply the Method Set. You will see the UV spectra and the MS spectra from ACQUITY QDa 2 scan in the Mass Analysis window (figure 4).

Empower Tips: View the UV spectra and the MS spectra | Figure 4

Figure 4

 

It’s that easy!

Final notes:

  • This procedure can be followed using the QuickStart or Pro interface.
  • ACQUITY QDa Mass Detector is compatible with the Alliance HPLC system.

 

Please rate this Empower Tip of the Week


Next week in Empower Tip #116– We will continue this series on working with ACQUITY QDa MS data in Empower (Part 16).

 