Welcome back to Get Empowered! In the last Empower tip-of-the-week post for Empower Chromatography Data Software, we learned why it is important to align chromatograms when collecting data from both a PDA and a QDa mass detector (Tip #105).

Working with ACQUITY QDa MS data in Empower:

  • #100: How to view the base mass and mass spectra for the peaks in a chromatogram
  • #101: How to work with 2D ACQUITY QDa MS data
  • #103: Viewing different spectral views in the Mass Analysis window
  • #104: Extracting chromatograms and spectra while working in the Mass Analysis window
  • #105:  Why it is important to align chromatograms when collecting data from both a PDA and a QDa mass detector

This week, we are going to learn how to optimize sensitivity by extracting a single mass rather than working with a TIC plot (Part 6). This is somewhat analogous to extracting a single wavelength rather than working with a MaxPlot when working with the PDA data.

Let’s get started.

Step 1

We start by bringing a 3D MS Scan into review and applying the appropriate Method Set. The Processing Method is set up to integrate the main peaks and calculate the signal to noise ratio for each (figure 1). (Note: the signal to noise ratio for Peak 6 is 13.3).

Empower Tips: Bringing 3D MS Scan into review/ apply appropriate Method Set | Figure 1

Figure 1

Step 2

To optimize sensitivity for Peak 6, we can derive a single mass and add it to our Method Set. In this case, we derive 630.55, which is the base mass for this peak (figure 2). (Refer back to Tip#56 which explains how to derive a single mass).

Empower Tips: Derive a single mass / add it to our Method Set | Figure 2

Figure 2

Step 3

Alternatively, we could use the ‘Extract Choices’ drop down list, located in the lower toolbar. Select the ‘blank’ and then enter the desired mass. This will automatically be added to the channel table in the Method Set (figure 3).

Empower Tips: Go to 'Extract Choices’ drop down list / lower toolbar | Figure 3

Figure 3

 

Step 4

Integrate the chromatogram for the single mass. The signal to noise for the peak at this mass is now 334.2 (figure 4).

Empower Tips: Integrate the chromatogram for single mass | Figure 4

Figure 4

It’s that easy!

 

Final notes: 

  • You are not limited to deriving one mass, i.e., you can derive multiple masses so that when you process with the Method Set all masses will be extracted and processed.
  • This procedure can be followed using the QuickStart or Pro interface.
  • ACQUITY QDa Mass Detector is compatible with the Alliance HPLC system.

 

Please rate this Empower Tip of the Week


Next week in Empower Tip #107– We will continue this series on working with ACQUITY QDa MS data in Empower (Part 7).

More resources

 